Thermoelectric Evaluation System
Atmospheric Thermoelectric Module Evaluation System
F-PEM
ULVAC EQUIPMENT SALES

For more information on ULVAC products, purchase, and/or consideration, click here.
Features
- The output of characteristics of the module can be evaluated by continuous loading at a high temperature in air.
- Repeated measurement of Pmax is possible at constant intervals while continuing to apply load for a long time.
- Measurement can be performed while applying a constant load in accordance with the actual module built-in environment.
Applications
- Evaluation of maximum power generation and heat flow of thermoelectric modules in air and under load
- Evaluation of conversion efficiency of module calculated from maximum power generation and heat flow
- Evaluation of durability of thermoelectric modules by long-hours measurement
Specifications
Measurement properties | Power generation, Heat flow, Conversion efficiency |
---|---|
Temperature range | Room temperature to 600℃ (heater setting value) |
Sample size | 40mm-square (Standard) |
Measurement atmosphere | In air |
Utility
Outside dimensions | W600 x D600 x H1700 (mm) |
---|---|
Weight | Approx. 120kg |
Power | Single phase AC200V, 5kW (main body) Single phase AC100V, 1kW (PC) |
Sample system chart
I : Current load
V : Voltage
C : Heat capacity of water
v : Flow rate of water
Tout : Outlet temp. of water
Tin : Inlet temp. of water
Power generation(P) = IV
Heat flow(Q) = Cv (Tout- Tin)
Conversio efficiency(η)=P / (P + Q)
Measurement example of oxide TEG
Pmax

Heat flow

Efficiency max