PEM-2|Power Generation Efficiency Characteristics Evaluation System|Thermoelectric Evaluation System|Heat Treatment And Thermal Properties|Products|ULVAC SHOWCASE

Thermoelectric Evaluation System
Power Generation Efficiency Characteristics Evaluation System
PEM-2

ULVAC EQUIPMENT SALES

For more information on ULVAC products, purchase, and/or consideration, click here.

Property evaluations of thermoelectric generator modules
By applying a maximum 500℃ temperature difference across a thermoelectric generator module, this system finds the thermoelectric conversion rate η by finding the power that is generated P when a one-dimensional heat flow Q is passed through the module.

Features

  • Uses an Infrared Gold Image Furnace with excellent temperature controllability to allow rapid module performance evaluations and endurance testing.
  • Capable of applying a maximum 500°C temperature difference across the top and bottom surface of the module.
  • Capable of measuring the amount of heat passed through the module.
  • Stably maintains the thermal resistance of the contact surface during heating with the air cylinder mechanism.
  • Measurement behavior can be configured with only software settings by temperature stability judgments, automatic variation of the load on the thermoelectric generator module, and automatic control of the measurement temperature.

Applications

  • Generation efficiency measurements of thermoelectric generator modules
  • Lifespan testing of modules by thermal cycle testing

Specifications

Measurement value Conversion efficiency, Power generation amount, Penetration heat amount
Measurement method One dimensional heat flow input method
Module size Square 30mm×5 to 30mm t (negotiable)
Heating surface temperature Max. Max. 800 ℃
Sample temperature gradient Approx. 300 ℃ in case of thermal conductivity of Stainless steel of 30 mm t
Pressure on contact surface 2MPa
Atmosphere In inert gas
Required floor space Approx. W3000×D1000 (mm)
Weight Approx. 300 kg
Power source Power : AC 200 V, 1 phase, 4 kVA, One location PC,
Pump : AC 100 V, 1 phase, 1 kVA Outlet two locations
Grounding Resistance 100 Ω max.
one location

Constitution

  • Sample heating assembly(including frame work)  1 set
  • Programmable temperature controller  1 set
  • Measurement circuit assembly  1 set
  • Vacuum evacuation assembly  1 set
  • Data processing unit  1 set
  • Constant temperature water circulator
    (for cooling lower part heat flow measurement block)  1 set

Utility

Required floor space Approx. W 3000 × D 1000 (mm)
Weight Approx. 300 kg
Power source Power : AC 200 V, 1 phase, 4 kVA, One location
PC, Pump : AC 100 V, 1 phase, 1 kVA
Outlet two locations
Grounding Resistance 100 Ω max.
one location

Typical example of commercially available module

img_pem-201.pngPower evaluation by temperature and load current

img_pem-202.pngEvaluation of temperature and maximum conversion efficiency

Inquiry about this product

1. I have read and agree to the Privacy Policy

Agree 

2. Enter the product name or model number.*

3. Select the status of the inquiry product.*

4. Request for *

5.  If you selected "I would like to consult about a problem or malfunction" in step 4, please enter the product number or serial number.

6. Comment *

​​​​​​​

Contact Information

Remember Me
​​​​​​​
​​​​​​​

The ULVAC email distribution office sends information about vacuum, such as product information, product maintenance, and web seminars. Click "Allow" if you wish.

Email delivery from us

This website use cookies to obtain and use access data to understand the convenience and usage of customers. If you agree to use cookies, click "I Accept".
[Privacy Plicy] [Cookie Policy]

I Accept