A series corresponding to special specifications for the properties of various thermoelectric materials and thin films.
A higher level evaluation for various kinds of materials has been required along with the development of thermoelectric materials. This system "Seebeck Coefficient / Electric Resistance. Measuring System ZEM-5 series" is specialized in the features of various kinds of materials such as high temperature, high resistance and thin film, etc. and a performance evaluation system to correspond to various kinds of needs.

Features

  • A series corresponding to special specifications for the characteristics of various thermoelectric materials and thin films
  • A type C thermocouple is used for the temperature detection sensor, making this system optimal for evaluating Si-based thermoelectric materials (SiGe, MgSi, others) (model HT)
  • V/I plot that self-diagnoses the ohmic contact is standard
  • Measurement of high temperature up to 1200°C (model HT)
  • Measurement of high resistance of maximum 10 MΩ (model HR)
  • Measurement of materials formed as a film on substrates (model TF)
  • Measurement of low temperature from -150°C to 200°C (model LT)

Applications

  • Evaluate the thermoelectric properties of a wide variety of materials including semiconductors, ceramics, and metals

Specifications

Model ZEM-5HT ZEM-5HR ZEM-5LT ZEM-5TF
Features High temperature High resistance Low and middle temperature Thin film
Temperature Range 100℃~1200℃ 100℃~800℃ -150℃~200℃ 50℃~500℃
Sample Size 2 to 4 mm square or Φ2 to 4 mm x 3 to 15 mm length Deposited substrate:2 to 4 mm
width x 0.4 to 1.2 mm
thickness mm x 20 mm length
Thin film thickness: nm scale or thicker
*An insulation layer is required between the sample film and the substrate
Thermocouple C R K R
Measurement accuracy Seebeck coefficient: ±7%
Electric resistivity: ±7% (less than 5mm long sample is not eligible.)

Patent and Standard

  • Thermoelectric power JIS R 1650-1
  • Resistivity JIS R 1650-2

Constitution

  • Sample measurement, control and measurement assy. 1 set
  • Data acquisition. 1 set
  • Atmosphere controller. 1 set

Utility(HT type)

Power Main body AC200V, 40A 1 location
PC AC100V, 10A 2 locations
Cooling water Water pressure 0.15 MPa Flow rate 5L/min
Atmosphere gas Low pressure He gas 0.2 MPa (Low pressure and static atmosphere)
Space requirement Approx.750mmW × 900mmD (Installed on table)

Seebeck coefficient calculation method

img_zem-501.pngSeebeck coefficient measurement shall be done by changing 3 temperature gradients and dividing dark electromotive force.

Setting and measurement screen

img_zem-502.pngTemperature setting screen

img_zem-503.pngV/I plot screen

該非判定結果報告書 Download

SDS Download

Accessories and Consumables

Instruction Manuals and Specifications Download

Applications Download

Technical drawings Download

Technical Data Download

This website use cookies to obtain and use access data to understand the convenience and usage of customers. If you agree to use cookies, click "I Accept".
[Privacy Plicy] [Cookie Policy]

I Accept