AFFINIX QN Pro(QCM2008-PRKIT)
ULVAC
For more information on ULVAC products, purchase, and/or consideration, click here.
Also, the built-in analysis mode enables the evaluation of physical properties such as viscoelastic analysis of adsorbates and calculation of mass and mucous membranes.
Specifications
| Item | AFFINIX QN Pro |
|---|---|
| Model | QCM2008-PRKIT |
| Measurement Principle・System | Quartz Crystal Microbalance(QCM) ・ Admittance Analysis Methods |
| Measurement Frequency・Sensiticity | 27MHz(Fundamental) ・ 30pg/Hz 81MHz(Overtone) ・ 10pg/Hz |
| Dynamic Range | 100pg~10μg |
| Oscillation Stability | Noise ≦ 2 Hz(27MHz, Distilled water, 25℃) |
| Average Drift | ≦ 1 Hz/min(27MHz, Distilled water, 25℃, at stable state) |
| Injection System | Manual, Multiple injection is possible |
| Min. Injection Volume | 0.1μL~ (dependent on micropipette to be used) |
| Channel Number | 1ch |
| Cup Volume | 400~550μL |
| Sensor Type | Sensor Cup(specific for QN Pro) |
| Stir Method | Magnetic Stirrer |
| Setting Temperature | 10.0~60.0℃ |
| System Software | Measuring software:『AFFINIX QN Pro』 (Since it is pre-installed in an attached PC, there is no installer disc) Data viewer software:『AFFINIX QN Pro Viewer』 (The installation system in saved at PC and additional disc) |
| Date Processing Part | Windows Laptop PC |
| Communication System | Ethernet Virtual serial PClcard |
| Output parameters | Fs,F2,Fw,Q,D,R,(C0,C1,L1) |
| Operating Environment | 20~30℃(Recomended) |
| Body Size | 140W×300D×220H(mm) (AFFINIX QN main body ecvept for analyzing unit) |
| Analysing Unit size | 71W×360D×144H(mm) |
| Weight | 9.0kg |
| Power Supply | AC 100~240V、3A (Body:1A, Analyzing Unit:2A) |
| Optional accessory | Data Analysis Software:『AQUA』 |
| STD accessories(Quantity) | ◆AFFINIX QN Pro Base Equipment(1) ◆Communication / Measurement Kit(1)
|