Optimal for process management of sputtering equipment (monitoring of residual moisture, impurities, and leaks)
Sputtering Process Model:CGM2-051 / 052
ULVAC

For more information on ULVAC products, purchase, and/or consideration, click here.
The Qulee offers high accuracy and resolution even at high pressure (1Pa or less).
Features
- Ideal for process monitoring of various sputtering sysytem
- Monitoring the Air Leak
- Management of residual moisture
- No differential pumping system required (less than 1 Pa at 7.5 x 10-3 Torr/1 x 10-2 mbar)
- Built-in display: PC-free
- Highly sensitive leak test possible (CGM2-052).More leak testing.
- Various leak tests available (Helium leak test, air leak test, assembly test)
- Total pressure measurement (ionization gauge) possible
- Includes a Quick Install Package system (QIP). Automatic measurement package (see option catalog for details)
- Qulee QCS is Included : This software is included and compatible with (Windows 8/10/11)
- Conforms with CE
Applications
- Process monitoring for PV, FPD, and semiconductor system
- Residual gas analysis in PVD system
- Impurities (H2O, etc.) control in PVD system
- For leak testing
Specifications
Model | CGM2-051 | CGM2-052 | CGM2-101 | CGM2-102 |
Mass filter type | Quadrupole | |||
Mass range | 1 to 50 amu | 1 to 100 amu | ||
Resolution | M/ΔM=1M (10% P.H.) | |||
Detector type | Faraday cup (FC) | EM Tube (SEM) / Faraday cup (FC) |
Faraday cup (FC) | EM Tube (SEM) / Faraday cup (FC) |
Sensitivity | 0.1uA/Pa | SEM: 0.1mA/Pa | 0.1uA/Pa | SEM: 0.1mA/Pa |
13.3 uA/Torr | SEM: 13.3 mA/Torr | 13.3 uA/Torr | SEM: 13.3 mA/Torr | |
10 uA/mbar | SEM: 10 mA/mbar | 10 uA/mbar | SEM: 10 mA/mbar | |
Minimum detectable partial pressure |
1×10-7 Pa | SEM: 1×10-10 Pa | 1×10-7 Pa | SEM: 1×10-10 Pa |
7.5×10-10 Torr | SEM: 7.5×10-13 Torr | 7.5×10-10 Torr | SEM: 7.5×10-13 Torr | |
1×10-9 mbar | SEM: 1×10-12 mbar | 1×10-9 mbar | SEM: 1×10-12 mbar | |
Maximum operating pressure |
FC: 2 Pa | FC: 2Pa, SEM: 1×10-2Pa |
FC: 2 Pa | FC: 2 Pa, SEM: 1×10-2Pa |
FC: 1.5×10-2 Torr | FC: 1.5×10-2 SEM: 7.5×10-5 Torr |
FC: 1.5×10-2 Torr | FC: 1.5×10-2 SEM: 7.5×10-5 Torr |
|
FC: 2×10-2 mbar | FC: 2×10-2 SEM: 1×10-4 mbar |
FC: 2×10-2 mbar | FC: 2×10-2 SEM: 1×10-4 mbar |
|
Linearity | 1Pa | |||
7.5×10-3 Torr | ||||
1×10-2 mbar | ||||
Total pressure function | Capable | |||
Degas function | Electron bombard type 300V, 5mA | |||
One Click function | Capable / He / H2O / N2 / O2 / Any gas | |||
One Click leak test function | Capable | |||
Power input | DC24V ±10% 50W | |||
Maximum sensor operating temperature |
120℃ (248℉) | |||
Maximum sensor operating temperature |
250℃ (482℉) | |||
Operating temperature range |
10 to 40℃ (50 to 104F) | |||
Operating humidity range |
15 to 80 % (not condensing) | |||
Communication interface | Ethernet | |||
Software | Qulee QCS Ver. 4.2 later (Windows 8/10/11) | |||
Applicable standard | CE |