Model:CGM2-101 / 102|Sputtering Process|Gas Analyzer(Process Gas Monitor)|Products|ULVAC SHOWCASE

Optimal for process management of sputtering equipment (monitoring of residual moisture, impurities, leaks, hydrocarbons, siloxanes, and garden-related compounds)

Sputtering Process Model:CGM2-101 / 102

ULVAC

For more information on ULVAC products, purchase, and/or consideration, click here.

New Product


The "Qulee" (pronounced "KLEE") is ULVAC's latest model for residual gas analysis and gas monitoring during the process. Feedback from equipment engineers on each production line has been incorporated into the new product design in pursuit of simplicity.
Measurements can be made with high accuracy at the process pressure of the sputtering equipment (less than 1 Pa) without using a differential exhaust system.

Features

  • Ideal for process monitoring of various sputtering sysytem
  • Monitoring the Air Leak
  • Management of residual moisture
  • Cleaning fluid residue
  • Galden® and other gas leaks
  • Monitoring of residual hydrocarbons
  • No differential pumping system required (less than 1 Pa at 7.5 x 10-3 Torr/1 x 10-2 mbar)
  • Built-in display: PC-free
  • Highly sensitive leak test possible (CGM2-102).More leak testing.
  • Various leak tests available (Helium leak test, air leak test, assembly test)
  • Total pressure measurement (ionization gauge) possible
  • Includes a Quick Install Package system (QIP). Automatic measurement package (see option catalog for details)
  • Qulee QCS is Included : This software is included and compatible with (Windows 8/10/11)
  • Conforms with CE

Applications

  • Ideal for residual gas analysis, process monitoring and impurity (H2O) management of sputtering equipment
  • For leak testing of sputtering equipment.

Specifications

Model CGM2-051 CGM2-052 CGM2-101 CGM2-102
Mass filter type Quadrupole
Mass range 1 to 50 amu 1 to 100 amu
Resolution M/ΔM=1M (10% P.H.)
Detector type Faraday cup (FC) EM Tube (SEM) /
Faraday cup (FC)
Faraday cup (FC) EM Tube (SEM) /
Faraday cup (FC)
Sensitivity 0.1uA/Pa SEM: 0.1mA/Pa 0.1uA/Pa SEM: 0.1mA/Pa
13.3 uA/Torr SEM: 13.3 mA/Torr 13.3 uA/Torr SEM: 13.3 mA/Torr
10 uA/mbar SEM: 10 mA/mbar 10 uA/mbar SEM: 10 mA/mbar
Minimum detectable
partial pressure
1×10-7 Pa SEM: 1×10-10 Pa 1×10-7 Pa SEM: 1×10-10 Pa
7.5×10-10 Torr SEM: 7.5×10-13 Torr 7.5×10-10 Torr SEM: 7.5×10-13 Torr
1×10-9 mbar SEM: 1×10-12 mbar 1×10-9 mbar SEM: 1×10-12 mbar
Maximum operating
pressure
FC: 2 Pa FC: 2Pa,
SEM: 1×10-2Pa
FC: 2 Pa FC: 2 Pa,
SEM: 1×10-2Pa
FC: 1.5×10-2 Torr FC: 1.5×10-2
SEM: 7.5×10-5 Torr
FC: 1.5×10-2 Torr FC: 1.5×10-2
SEM: 7.5×10-5 Torr
FC: 2×10-2 mbar FC: 2×10-2
SEM: 1×10-4 mbar
FC: 2×10-2 mbar FC: 2×10-2
SEM: 1×10-4 mbar
Linearity 1Pa
7.5×10-3 Torr
1×10-2 mbar
Total pressure function Capable
Degas function Electron bombard type 300V, 5mA
One Click function Capable / He / H2O / N2 / O2 / Any gas
One Click leak test function Capable
Power input DC24V ±10% 50W
Maximum sensor
operating temperature
120℃ (248℉)
Maximum sensor
operating temperature
250℃ (482℉)
Operating temperature
range
10 to 40℃ (50 to 104F)
Operating humidity
range
15 to 80 % (not condensing)
Communication interface Ethernet
Software Qulee QCS Ver. 4.2 later (Windows 8/10/11)
Applicable standard CE

Inquiry about this product

1. I have read and agree to the Privacy Policy

Agree 

2. Enter the product name or model number.*

3. Select the status of the inquiry product.*

4. Request for *

5.  If you selected "I would like to consult about a problem or malfunction" in step 4, please enter the product number or serial number.

6. Comment *

​​​​​​​

Contact Information

Remember Me
​​​​​​​
​​​​​​​

The ULVAC email distribution office sends information about vacuum, such as product information, product maintenance, and web seminars. Click "Allow" if you wish.

Email delivery from us

This website use cookies to obtain and use access data to understand the convenience and usage of customers. If you agree to use cookies, click "I Accept".
[Privacy Plicy] [Cookie Policy]

I Accept