Model:GI-D7(WIT)|Hot Cathode Gauge|Vacuum Gauge|Products|ULVAC SHOWCASE

Hot cathode ionization gauge that uses glass tube sensor heads (triode type)

Hot Cathode Gauge Model:GI-D7(WIT)

ULVAC

For more information on ULVAC products, purchase, and/or consideration, click here.

This model is hot cathode type ionization vacuum gauge, capable of measuring pressure in the 6.7 x 10-1 to 10-6 Pa (5 x 10-3 to 9.75 x10-9 Torr, 6.7 x 10-3 to 1.3 x 10-8 mbar)(Glass gauge) range using triode tube or B-A type sensor heads.

Features

  • 2 types vacuum tube, metal and glass: Sensor head selected depending on the application (bare type also available)
  • Sensitivity setup can be changed for different gases. Allowing even more accurate pressure measurement.
  • Ideal for fixing the measurement range
  • 2 set points output
  • Switch 1/2 filament, sensitivity setting, range hold function, Degas function
  • RS232C communication

Applications

  • Vacuum heat treatment furnace
  • Vacuum brazing furnace
  • Vacuum evaporation system(diffusion pump type)
  • Process control for furnaces and other industrial equipment

Specifications

Product name Glass Ionization Vacuum Gauge
Model model GI-D7
Measurable pressure range WIT:1.30×10-5 to 6.7×10-1 Pa
WIB:1.30×10-6 to 2.3×10-2 Pa
Accuracy ±15%
Pressure indication Didital indication 2 digit mantissa, 1 digit exponent
Compatible sensor head WIT-G1, WIT-G2, WIT-G3, WIB-G5, WIB-N3
Sampling time 100ms
DEGAS Direct electrical heating of grid AC power 1.5VA
Functions Range hold function, Gas sensitivity setting, Filament selection
Measurement value output linear output 0 to 10 VDC, Pseudo-logarithmic output 0-10VDC(1V linear signal in each digit), Log output
Control input REMOTE/LOCAL, Filament ON/OFF, Filament 1/2, DEGAS ON/OFF, Range hold
Control output Analog output(0-10V), BCD output, Filament ON/OFF, Emission current setting, Degassing ON/OFF
Set point(contact point capacity) 2 point(AC125Vmax, 0.5Amax / DC24Vmax, 1.0Amax)
Serial communication spec RS232C
Operating temperature range 10 to 40℃(50 to 104℉)
Operating humidity range 15 to 18%(not condensing)
Power supply voltage AC100±10V 50/60Hz, 120VA
Fuse BS. SEMKO 3.15A (2pcs)
External dimensions W×D×H 240mm×380mm×99mm
Weight 5.7kg
Standard accessories Power cord 3P Plug with ground
Connector For external input / output
Quick manual 1 pc
Options Sensor cable 4, 10, 15, 20, 25, 30m
Gauge head Triode type glass gauge:WIT-G1, WIT-G2, WIT-G3
B-A type glass gauge:WIB-G5
B-A tupe nude gauge:WIB-N3

Inquiry about this product

1. I have read and agree to the Privacy Policy

Agree 

2. Enter the product name or model number.*

3. Select the status of the inquiry product.*

4. Request for *

5.  If you selected "I would like to consult about a problem or malfunction" in step 4, please enter the product number or serial number.

6. Comment *

​​​​​​​

Contact Information

Remember Me
​​​​​​​
​​​​​​​

The ULVAC email distribution office sends information about vacuum, such as product information, product maintenance, and web seminars. Click "Allow" if you wish.

Email delivery from us

This website use cookies to obtain and use access data to understand the convenience and usage of customers. If you agree to use cookies, click "I Accept".
[Privacy Plicy] [Cookie Policy]

I Accept