Simultaneously measures the Seebeck coefficient and electric conductivity of thermoelectric materials.
Based on requests from both domestic and international academic societies and customers, this system features our easy to use, high-precision measurement techniques. Complies with JIS standards for thermoelectric material Seebeck coefficient measurement equipment.

Features

  • A Infrared Gold Image Furnace with excellent temperature controllability and a micro-heater for temperature difference control
  • The measurements are controlled by a computer and enable to perform the measurements at specified temperatures on each emperature difference which eliminates dark electromotive force and allows automatic measurements
  • Ohmic contact automatic check function (V-I plot) as a standard function
  • Available to measure a thin film with an original adapter as an option.
  • Customizable to high resistance type

Applications

  • Evaluate the thermoelectric properties of a wide variety of materials including semiconductors, ceramics, and metals

Specifications

Model ZEM-3L ZEM-3LW ZEM-3M8 ZEM-3M10
Temperature Range -80℃ to 100℃ -100℃ to 200℃ RT, 50℃ to 800℃ RT, 50℃ to 1000℃
Measurement Properties Seebeck coefficient, electric resistivity
Sample Size 2 to 4 mm square or Φ x 5 to 22 mm length
Measurement Atomosphere Low-pressure He gas
Optional Cooling water circulator
Thermoelectric generation simulation software
Thin film attachment
Measurements in various atmospheres (negotialble)

A patent and a standard

  • Thermoelectric power JIS R 1650-1
  • Resistivity JIS R 1650-2

Measurement principle

img_zem-301-en.png

A prism or cylindrical sample is set in a vertical position between the upper and lower blocks in the heating furnace. While the sample is heated, and held, at a specified temperature, it is heated by the heater in the lower block to provide a temperature gradient. Seebeck coefficient is measured by measuring the upper and lower temperatures T1 and T2 with the thermocouples pressed against the side of the sample, followed by measurement of thermal electromotive force dE between the same wires on one side of the thermocouple. Electric resistance is measured by the dc four-terminal method, in which a constant current I is applied to both ends of the sample to measure and determine voltage drop dV between the same wires of the thermocouple by subtracting the thermo-electromotive force between leads.

Measurement Example of P Type Si80Ge20 Sintered Compact

img_zem-302.png

img_zem-303.png

該非判定結果報告書 Download

SDS Download

Accessories and Consumables

Instruction Manuals and Specifications Download

Applications Download

Technical drawings Download

Technical Data Download