Model: Gv200|Magnetic Coupling Type|Oil Rotary Vacuum Pump|Rough&Middle Vacuum Pump|Products|ULVAC SHOWCASE

Oil leak-free power transfer section

Oil Rotary Vacuum Pump
Magnetic Coupling Type
Model: Gv200

ULVAC

For more information on ULVAC products, purchase, and/or consideration, click here.

Double-stage exhaust system / Including backflow prevention mechanism / Gas ballast mechanism / Built-in thermal protector / ON/OFF switch / Input terminal (cord not included)

User advantages

  • Quiet operation suitable for offices and laboratories in quiet environments.
  • No need to handle or clean oil leaks from the vacuum pump.
  • Prevents oil backflow into equipment, ensuring contamination-free operation.

Specification

Model Unit 50Hz 60Hz
Actual Pumping Speed L/min 200 240
Ultimate Pressure*¹ (G.V.Close) Pa 5.0 x 10‐¹
Ultimate Pressure (G.V. open) Pa 5
Motor Output W 550
Recommended oil   SMR-100
Oil capacity mL 1,000
Weight kg 31
Inlet port diameter mm KF-25
Ambient temperature 7 - 40
Overall dimensions mm 166(W) x 490(L) x 241.5(H)
RoHS applied  
*1 The ultimate pressure is measured by a Pirani gauge. For pumps with an ultimate pressure of 1.3 Pa or lower, measurements taken with a McLeod gauge may show values approximately one order of magnitude lower.
* The power plug and cord are not included.

Power Supply/Voltage

Voltage Certificate
AC1φ 100-120 / 200-240V TUV, cTUVus, ATEX(category 3)

Remarks

1. Pump oil deteriorates during operation, which can lead to reduced performance. Please regularly inspect the oil level and contamination, and refill or replace the oil as needed.
2. The backflow prevention valve mechanism prevents pump oil from flowing back into the vacuum chamber during emergencies such as power outages.
When the pump will be stopped for an extended period, please return the vacuum chamber to atmospheric pressure.
3. Oil mist may be generated from the exhaust port during operation, especially near atmospheric pressure.
4. Continuous operation at a high intake pressure of 5,000Pa or above is not suitable.
5. Operating the pump with the exhaust port blocked or with devices that obstruct gas flow on the exhaust side may cause rupture.
6. The motor's overload protection device is of the automatic reset type.
7. If the overload protection device activates during pump operation, the pump will automatically restart once the motor winding temperature decreases.
To prevent accidents, immediately unplug the power cord if the overload protection device activates.

Outline Drawing

Gv200_drawing.png


Pumping Speed Curves

Gv135_200_speedcurves.png