Applications of Process gas monitor.|Technology|Solution|ULVAC SHOWCASE

Technology

(1)Application Data Deposition Equipment Residual Gas Data.

gas01_1.jpg

Atmospheric pressure components. N2:78% O2:21% Ar:0.9%

・The normal high vacuum state has mainly H2O and H2.

・The existence of leak can be judged by the ionic strength of N2(28) and O2(32) and their ratios.

・Contamination of oil components can be judged by M/e=39,41,43,55,57(CH).

(2)Application Data PVD Process Gas DataSemi

gas02_1.jpg

(3)Application Data PVD Process Gas Data FPD

At the time of Ar Gas introduction (2).(When introducing Ar 400 sccm, O2 is introduced 2 → 4 → 6 → 8 → 10sccm.)

gas03_1.jpg

(4)Application Data  PVD Process Gas Data semi

gas04_1.jpg

(5)Application Data  Vacuum Drying Furnace ・Process Data

gas05_1.jpg

Technology

Solution

This website use cookies to obtain and use access data to understand the convenience and usage of customers. If you agree to use cookies, click "I Accept".
[Privacy Plicy] [Cookie Policy]

I Accept